Maintaining the MSD
99
Chemical residue is another potential danger
Only a small portion of your sample is ionized by the ion source. The
majority of any sample passes through the ion source without being ionized.
It is pumped away by the vacuum system. As a result, the exhaust from the
foreline pump will contain traces of the carrier gas and your samples. It will
also contain tiny droplets of foreline pump oil.
An oil trap is supplied with the foreline pump. This trap stops
only
pump
oil droplets. It
does not
trap any other chemicals. If you are using toxic
solvents or analyzing toxic chemicals, do not use the oil trap. Instead, install
a hose to take the exhaust from the foreline pump outdoors or into a fume
hood vented to the outdoors. Be sure to comply with your local air quality
regulations.
The fluids in the diffusion pump and foreline pump also collect traces of the
samples being analyzed. All used pump fluid should be considered hazard-
ous and handled accordingly. Dispose of used fluid correctly, as specified
by your local regulations.
: $ 5 1 , 1 *
7KHRLOWUDSVWRSVRQO\IRUHOLQHSXPSRLO,WGRHVQRWWUDSRUILOWHURXWWR[LFFKHPLFDOV,I\RX
DUHXVLQJWR[LFVROYHQWVRUDQDO\]LQJWR[LFFKHPLFDOVUHPRYHWKHRLOWUDS'RQRWXVHWKHWUDS
LI\RXKDYHD&,06',QVWDOODKRVHWRWDNHWKHIRUHOLQHSXPSH[KDXVWRXWVLGHRUWRDIXPH
KRRG
: $ 5 1 , 1 *
:KHQUHSODFLQJSXPSIOXLGXVHDSSURSULDWHFKHPLFDOUHVLVWDQWJORYHVDQGVDIHW\JODVVHV
$YRLGDOOFRQWDFWZLWKWKHIOXLG
Summary of Contents for 5973
Page 1: ...H Hardware Manual HP 5973 Mass Selective Detector ...
Page 3: ...HP 5973 Mass Selective Detector Hardware Manual ...
Page 25: ...1 Installing GC Columns To condition a capillary column 25 ...
Page 30: ...30 ...
Page 37: ...2 Operating the MSD To view MSD analyzer temperature and vacuum status 37 ...
Page 39: ...2 Operating the MSD To set monitors for MSD temperature and vacuum status 39 ...
Page 41: ...2 Operating the MSD To set the MSD analyzer temperatures 41 ...
Page 43: ...2 Operating the MSD To set the interface temperature from the ChemStation 43 ...
Page 51: ...2 Operating the MSD To remove the MSD covers 51 Upper cover Lower cover ...
Page 92: ...92 ...
Page 105: ...4 Maintaining the MSD To drain the foreline pump 105 Fill cap Drain plug Pump motor ...
Page 113: ...4 Maintaining the MSD To separate the MSD from the GC 113 ...
Page 121: ...4 Maintaining the MSD To reconnect the MSD to the GC 121 ...
Page 141: ...4 Maintaining the MSD To lubricate the side plate O ring 141 Side plate O ring ...
Page 159: ...4 Maintaining the MSD To remove a filament 159 Filament 2 Filament 1 ...
Page 161: ...4 Maintaining the MSD To reinstall a filament 161 Filament 2 Filament 1 ...
Page 210: ...210 ...
Page 214: ...214 ...
Page 232: ...232 7 Analyzer Analyzer heaters and radiators ...
Page 280: ...280 ...
Page 294: ...294 Index ...
Page 296: ...294 Chapter ...
Page 297: ...H Manual Part Number G1099 90027 Copyright 1998 Hewlett Packard Printed in USA 11 98 ...