3
Troubleshooting the MSD
Pressure symptoms
79
Pressure symptoms
This section describes unusual pressure readings and their possible causes. The
symptoms in this section are based on typical pressures. At typical column flow
rates (0.1 - 2.0 ml/minute), the foreline pressure will be approximately 20 to
100 mTorr. The vacuum manifold pressure will be approximately 1
×
10
-6
to
1.4
×
10
-4
Torr. These pressures can vary widely from instrument to instrument so
it is very important that you are familiar with the pressures that are typical for
your instrument at given carrier gas flows.
The foreline pressures listed can only be measured on diffusion pump-equipped
systems. Turbomolecular pumps are controlled according to their speed and do
not have foreline pressure gauges. The vacuum manifold pressures can only be
measured if your system is equipped with the optional gauge controller.
Foreline pressure is too high
If the pressure you observe is above 100 mTorr, or if the pressure you observe for a
given column flow has increased over time, check the following:
•
Column (carrier gas) flow is too high
•
Air leak (usually the sideplate is not pushed in or vent valve is open)
•
Foreline pump oil level is low or oil is contaminated
•
Foreline hose is constricted
•
Foreline gauge is not working correctly
•
Foreline pump is not working correctly
Vacuum manifold pressure is too high
If the pressure you observe is above 1.4
×
10
-4
Torr, or if the pressure you observe
for a given column flow has increased over time, check the following:
•
Column (carrier gas) flow is too high
•
Air leak
•
Foreline pump is not working correctly (see
Foreline pressure is too high
)
•
Diffusion pump fluid level is low or fluid is contaminated
•
Turbomolecular pump is not working correctly
Summary of Contents for 5973
Page 1: ...H Hardware Manual HP 5973 Mass Selective Detector ...
Page 3: ...HP 5973 Mass Selective Detector Hardware Manual ...
Page 25: ...1 Installing GC Columns To condition a capillary column 25 ...
Page 30: ...30 ...
Page 37: ...2 Operating the MSD To view MSD analyzer temperature and vacuum status 37 ...
Page 39: ...2 Operating the MSD To set monitors for MSD temperature and vacuum status 39 ...
Page 41: ...2 Operating the MSD To set the MSD analyzer temperatures 41 ...
Page 43: ...2 Operating the MSD To set the interface temperature from the ChemStation 43 ...
Page 51: ...2 Operating the MSD To remove the MSD covers 51 Upper cover Lower cover ...
Page 92: ...92 ...
Page 105: ...4 Maintaining the MSD To drain the foreline pump 105 Fill cap Drain plug Pump motor ...
Page 113: ...4 Maintaining the MSD To separate the MSD from the GC 113 ...
Page 121: ...4 Maintaining the MSD To reconnect the MSD to the GC 121 ...
Page 141: ...4 Maintaining the MSD To lubricate the side plate O ring 141 Side plate O ring ...
Page 159: ...4 Maintaining the MSD To remove a filament 159 Filament 2 Filament 1 ...
Page 161: ...4 Maintaining the MSD To reinstall a filament 161 Filament 2 Filament 1 ...
Page 210: ...210 ...
Page 214: ...214 ...
Page 232: ...232 7 Analyzer Analyzer heaters and radiators ...
Page 280: ...280 ...
Page 294: ...294 Index ...
Page 296: ...294 Chapter ...
Page 297: ...H Manual Part Number G1099 90027 Copyright 1998 Hewlett Packard Printed in USA 11 98 ...