Revision 1.0
SMV800 Modbus User's Guide
Page 205
Information
Miscellaneous
DB Corrupt
NVM copy of
Miscellaneous
Configuration
database block found
corrupt.
Power cycle the
device. If the problem
persists, replace the
transmitter
Bit - 0
123
Totalizer Value
DB Corrupt
NVM copy of totalizer
value block is found
corrupt
Power cycle the
device. If the problem
persists, replace the
transmitter
Bit -1
Flow Unit DB
Corrupt
NVM copy of totalizer
value block is found
corrupt
Power cycle the
device. If the problem
persists, replace the
transmitter
Bit - 2
Flow Parameter
DB Corrupt
NVM copy of
advance flow value
block is found corrupt
Power cycle the
device. If the problem
persists, replace the
transmitter
Bit - 3
Backup Totalizer
Value DB
Corrupt
NVM copy of totalizer
value back up block
is found corrupt
Power cycle the
device.
Bit - 4
User Variable DB
Corrupt
NVM copy of user
variables block is
found corrupt
Power cycle the
device. If the problem
persists reconfigure
user defined variables
LRV,URV if
required.If still
problem persists
replace transmitter.
Bit - 5
Summary of Contents for SMV800
Page 82: ...Page 70 SMV800 Modbus User s Guide Revision 1 0 ...
Page 95: ...Revision 1 0 SMV800 Modbus User s Guide Page 83 2 5 4 Review ...
Page 99: ...Revision 1 0 SMV800 Modbus User s Guide Page 87 ...
Page 100: ...Page 88 SMV800 Modbus User s Guide Revision 1 0 ...
Page 101: ...Revision 1 0 SMV800 Modbus User s Guide Page 89 ...
Page 102: ...Page 90 SMV800 Modbus User s Guide Revision 1 0 ...
Page 103: ...Revision 1 0 SMV800 Modbus User s Guide Page 91 Maintenance Menu ...
Page 104: ...Page 92 SMV800 Modbus User s Guide Revision 1 0 2 7 1 Advanced Diagnotics Menu ...
Page 105: ...Revision 1 0 SMV800 Modbus User s Guide Page 93 2 7 2 Monitoring Menu ...
Page 112: ...Page 100 SMV800 Modbus User s Guide Revision 1 0 ...
Page 116: ......