TECHNICAL DATA
8
89
SMARTLINE® RM60 (FMCW)
www.honeywell.com/ps
34-VF-25-31 iss.4 GLO December 15 US
Measurement accuracy
Resolution
1 mm / 0.04
¨
Repeatability
±
1 mm /
±
0.04
¨
Accuracy
±
3 mm /
±
0.12
¨
, when distance < 10 m / 33 ft;
±
0.03% of measured distance, when distance > 10 m / 33 ft
Reference conditions acc. to EN 60770
Reference conditions acc. to EN 60770
Reference conditions acc. to EN 60770
Reference conditions acc. to EN 60770
Temperature
+20
°
C
±
5
°
C / +70
°
F
±
10
°
F
Pressure
1013 mbara
±
20 mbar / 14.69 psia
±
0.29 psi
Relative air humidity
60%
±
15%
Target
Metal plate in an anechoic chamber
Operating conditions
Temperature
Temperature
Temperature
Temperature
Ambient temperature
-40
…
+80
°
C / -40
…
+175
°
F (according to the temperature limits of the gasket
material. Refer to "Materials" in this table.)
Ex: see supplementary operating instructions or approval certificates
Storage temperature
-40
…
+85
°
C / -40
…
+185
°
F
Flange temperature
Horn / Sheet metal horn antenna:
Horn / Sheet metal horn antenna:
Horn / Sheet metal horn antenna:
Horn / Sheet metal horn antenna:
Standard: -50
…
+150
°
C / -58
…
+300
°
F
Option: -50
…
+200
°
C / -58
…
+390
°
F
(the process connection temperature must agree with the temperature limits of the
gasket material. Refer to "Materials" in this table.)
Ex: see supplementary operating instructions or approval certificates
Drop antenna (PTFE):
Drop antenna (PTFE):
Drop antenna (PTFE):
Drop antenna (PTFE):
-50
…
+150
°
C / -58
…
+300
°
F (the process connection temperature must agree with
the temperature limits of the gasket material. Refer to "Materials" in this table.)
Ex: see supplementary operating instructions or approval certificates
Drop antenna (PP):
Drop antenna (PP):
Drop antenna (PP):
Drop antenna (PP):
-40
…
+100
°
C / -40
…
+210
°
F (the process connection temperature must agree with
the temperature limits of the gasket material. Refer to "Materials" in this table.)
Ex: see supplementary operating instructions or approval certificates
Pressure
Pressure
Pressure
Pressure
Operating pressure
Drop antenna (PP):
Drop antenna (PP):
Drop antenna (PP):
Drop antenna (PP):
-1
…
16 barg / -14.5
…
232 psig;
subject to process connection used and flange temperature
Drop antenna (PTFE):
Drop antenna (PTFE):
Drop antenna (PTFE):
Drop antenna (PTFE):
-1
…
40 barg / -14.5
…
580 psig;
subject to process connection used and flange temperature
Horn / Sheet metal horn antenna:
Horn / Sheet metal horn antenna:
Horn / Sheet metal horn antenna:
Horn / Sheet metal horn antenna:
Standard: -1
…
40 barg / -14.5
…
580 psig;
subject to process connection used and flange temperature
Other conditions
Other conditions
Other conditions
Other conditions
Dielectric constant (
ε
r)
≥
1.5
Vibration resistance
IEC 60068-2-6 and EN 50178 (10...57 Hz: 0.075 mm / 57...150 Hz:1g)
Protection category
IP66/67 equivalent to NEMA type 4X (housing) and type 6P (antenna)
Maximum rate of change
10 m/min / 33 ft/min
Summary of Contents for SmartLine RM60
Page 8: ...CONTENTS 8 www honeywell com ps 34 VF 25 31 iss 4 GLO December 15 US SMARTLINE RM60 FMCW ...
Page 76: ...6 OPERATION 76 SMARTLINE RM60 FMCW www honeywell com ps 34 VF 25 31 iss 4 GLO December 15 US ...
Page 86: ...7 SERVICE 86 SMARTLINE RM60 FMCW www honeywell com ps 34 VF 25 31 iss 4 GLO December 15 US ...
Page 126: ...10 APPENDIX 126 SMARTLINE RM60 FMCW www honeywell com ps 34 VF 25 31 iss 4 GLO December 15 US ...
Page 127: ...APPENDIX 10 127 SMARTLINE RM60 FMCW www honeywell com ps 34 VF 25 31 iss 4 GLO December 15 US ...