A-16
I/O Error ....
Hardware
or
Software
There was an error reading or writing a disk drive. The message should identify the
drive and say something about the nature of the error.
Action: If this is an "out of space" error, verify that there is room on the target drive.
Check the "Room for ..." message in the status window. Otherwise, run appropriate
diagnostics on the disk and controller; if the problem involves diskette (Drive A:), try a
different diskette.
Limit Exceeded: ......
Software
Any message that begins this way should be reported to Hologic Technical Support.
Include as much information as possible about what you were doing at the time.
Local motion active [E_LOCAL_MOTION_I]
Comment: An application has attempted to enable or disable local motion when one or
more motors were still moving. This is also the abort code reported by the state
machine when a protocol is aborted to start the local motion state machine.
Action: When moving the system using the operator control panel, wait until all
motion stops before starting an application that requires motor movement (e.g., the
scan program).
Machine serial number is not valid.
Action: QDR 4500 ID file C:\SQUID.DAT is not valid. Call Hologic Technical
Support.
Machine type in configuration files is inconsistent.
Action: QDR 4500 ID file C:\SQUID.DAT is not valid. Call Hologic Technical
Support.
Machine type is not valid.
Action: QDR 4500 ID file C:\SQUID.DAT is not valid. Call Hologic Technical
Support.
Memory allocation error.
Action: Verify that only Hologic software is installed on the system.
Message packet canceled [E_CANCELED_I]
Comment: The message packet was canceled before the response was received. This is
not an error but merely records the current state of a message packet. The driver
normally cancels all outstanding message packets when it goes off line.
Action: None.
ERROR MESSAGE
HARDWARE
/SOFTWARE
ERROR
LOCUS
COMMENT/ACTION
Summary of Contents for DISCOVERY A
Page 52: ...Discovery QDR Series Technical Manual 2 30...
Page 55: ...Discovery QDR Series Technical Manual 3 3 Figure 3 2 System Dimensions for Discovery C...
Page 56: ...Discovery QDR Series Technical Manual 3 4 Figure 3 3 System Dimensions for Discovery SL...
Page 144: ...Discovery QDR Series Technical Manual 4 22...
Page 188: ...Discovery QDR Series Technical Manual 5 22...
Page 228: ...Discovery QDR Series Technical Manual 6 16...
Page 248: ......