A-3
Error Message List
ERROR MESSAGE
HARDWARE
/SOFTWARE
ERROR
LOCUS
COMMENT/ACTION
Address in the response message did not match the address in the sent message. [E_ADDRESS_MISMATCH_I]
Action: This message should be reported to Hologic Technical Support. Include as
much information as possible about the scan.
Analysis aborted. d0= ___ below acceptable limit
Hardware
or
Software
d0 is found to be below zero, which is not possible if there are X-rays reaching the
detector.
Action: This message should be reported to Hologic Technical Support. Include as
much information as possible about the scan.
Arm and table are in danger of colliding! [E_COLLISION_IMMINENT_I]
Comment: The C/C has calculated that the C-arm and table are within 1/4".
Action: If the C-arm and table are really within this distance, the problem is with the
application (or operator) that moved them so close. If they are not, then either the
calibration information for the system is wrong or the mechanical stops for the table
and C-arm have moved. One or more of the motors may have stopped short of the
actual mechanical stop during calibration. Verify that the positive limit positions for
each motor correspond to the published specs for the model. To recover from this,
select the Utility option from the main menu and then select the emergency Motion
option from the sub-menu. By holding down the Enable button on the operator control
panel, you can use the keypad to move the table and C-arm away from each other.
Summary of Contents for DISCOVERY A
Page 52: ...Discovery QDR Series Technical Manual 2 30...
Page 55: ...Discovery QDR Series Technical Manual 3 3 Figure 3 2 System Dimensions for Discovery C...
Page 56: ...Discovery QDR Series Technical Manual 3 4 Figure 3 3 System Dimensions for Discovery SL...
Page 144: ...Discovery QDR Series Technical Manual 4 22...
Page 188: ...Discovery QDR Series Technical Manual 5 22...
Page 228: ...Discovery QDR Series Technical Manual 6 16...
Page 248: ......