Discovery QDR Series Technical Manual
6-2
The following suggestions apply to a Discovery system exhibiting a power problem:
Also...
6.5 Motion Problems
Motion problems are failures related to movement of the table and/or C-arm. In most (but
not all) cases, an error message accompanies Discovery motion problems. Therefore, it is
good troubleshooting practice to check the error log (C:\ERRLOG.DAT file) for
occurrences of motion related problems. The following suggestions apply to a Discovery
system that exhibits a motion problem.
Start by identifying the bad axis (if it is not obvious). Ask the operator for symptoms and
check the error log. See Figure 6-1 for possible motion directions.
If...
Check...
Refer to...
System "dead”
Main breaker left pedestal
Emergency Stop or Instrument On/Off
switch on Control Panel
Check Power Line Voltage
heading, Section 3
Check fuses
Main power suspect
Check AC input voltage
Power cable to the Torroid Assembly
Power Module heading,
Section 2
Suspected Power Module
problem
Check AC Voltage at Scanner DIN Rail
Power Module FRU s
heading, Section 5
Suspected DC power supply
problem
Check LED’s on Distribution Board for
all DC voltages (lit if present)
Check fuses
24V LED not lit
All tape switches and connectors (tape sw
reads 120W when open
Emergency Stop Switch and associated
circuitry
Computer is off
Hologic software running
(check error log)
Check fuses
Distribution board,
Figure 5-1
System does not start up
System properly configured (see “Before
Starting” on page 6-1
Before starting, make sure the software
configuration is compatible with the
scanner.
Emergency Stop switch in “On” (Out)
position
Console does not power up
Main circuit breaker, power strip
Problem...
Possible cause...
Corrective action...
Laser does not
turn on
Safety feature if system thinks C-arm
is at head of table
Restart in Service mode
From Service Utilities, run SQDRIVER
At CARM$$$$> prompt, type
LASER_SAFTY 0 <Enter>
Exit from SQDRIVER
Reboot the computer
Summary of Contents for DISCOVERY A
Page 52: ...Discovery QDR Series Technical Manual 2 30...
Page 55: ...Discovery QDR Series Technical Manual 3 3 Figure 3 2 System Dimensions for Discovery C...
Page 56: ...Discovery QDR Series Technical Manual 3 4 Figure 3 3 System Dimensions for Discovery SL...
Page 144: ...Discovery QDR Series Technical Manual 4 22...
Page 188: ...Discovery QDR Series Technical Manual 5 22...
Page 228: ...Discovery QDR Series Technical Manual 6 16...
Page 248: ......