Discovery QDR Series Technical Manual
1-13
1.3.2 Duty Cycle:
1.3.3 Leakage Technique Factors
The leakage technique factors for all models of Discovery are the same. It is the maximum
continuous current at the maximum peak potential. This is X-ray mode #3. Peak potential
100/140kVp (dual energy), current 10mA peak 25% duty cycle or 2.5mA average.
1.3.4 Minimum Beam Filtration
The minimum filtration permanently in the beam is 3.7mm Al equivalent @80kV.
1.3.5 Measured Half Value Layer (HVL) At Different Operat-
ing Potentials
1.3.6 Line Voltage and Maximum Line Current
1.3.7 Technique Factors for Maximum Line Current
Peak Potential 140kVp
Tube Current 10mA peak, 50% duty factor or 5mA average.
C, W and SL
IVA
12%
All Others
50%
A
Whole Body
100%
IVA
12%
All Others
50%
Measured operating potential
Measured Half Value Layer
Discovery All
80kV
3.7mm Al equivalent
100kV
5.0mm Al equivalent
140kV
6.5mm Al equivalent
Power
Requirements:
All
100VAC 16A 50/60Hz, Max apparent resistance = 0.32 ohm
120VAC 14A 50/60Hz, Max apparent resistance = 0.32 ohm
230VAC 8A 50/60Hz, Max apparent resistance = 1.28 ohm
Summary of Contents for DISCOVERY A
Page 52: ...Discovery QDR Series Technical Manual 2 30...
Page 55: ...Discovery QDR Series Technical Manual 3 3 Figure 3 2 System Dimensions for Discovery C...
Page 56: ...Discovery QDR Series Technical Manual 3 4 Figure 3 3 System Dimensions for Discovery SL...
Page 144: ...Discovery QDR Series Technical Manual 4 22...
Page 188: ...Discovery QDR Series Technical Manual 5 22...
Page 228: ...Discovery QDR Series Technical Manual 6 16...
Page 248: ......