Discovery QDR Series Technical Manual
3-10
Figure 3-7. Table X Drive
3.2.6 Remove Discovery A or SL Lower C-Arm Assembly (If
Necessary)
Note: It is not necessary to remove the C-arm if the hallway that the unit must go through
is at least 45" (114cm) wide (side-to-side inside dimension). Removing the C-arm
allows the unit to fit through a hallway 29" (74cm) wide.
The C-arm Carriage Assembly can be removed before taking the unit off the pallet.
Follow the procedure below to remove the C-arm:
1. Remove the 2 carriage-to-base shipping brackets (see Figure 3-8).
2. Remove the table locking bracket located on the left side of the scanner (see
3. Move the table forward, remove the C-Arm Interface board cover, and remove
the tank cover. Then move the table back.
4. Remove the X-Ray Controller Assembly (4 Phillips screws). See “X-Ray Con-
troller Assembly” on page 5-18 for detailed removal information.
5. Remove the Tank Assembly (leave Filter Drum Assembly attached). See “Tank
Assembly” on page 5-19 for detailed removal information.
6. Disconnect the cables and Nylatrac mounting hardware, and tape Nylatrac and
cables in the bottom of the scanner so that they are secure.
Stainless Steel
Cable Guard
X Drive Attachment Bracket
Front Endcap
Right Angle Bracket
Bearing Block
Bearing Block
Summary of Contents for DISCOVERY A
Page 52: ...Discovery QDR Series Technical Manual 2 30...
Page 55: ...Discovery QDR Series Technical Manual 3 3 Figure 3 2 System Dimensions for Discovery C...
Page 56: ...Discovery QDR Series Technical Manual 3 4 Figure 3 3 System Dimensions for Discovery SL...
Page 144: ...Discovery QDR Series Technical Manual 4 22...
Page 188: ...Discovery QDR Series Technical Manual 5 22...
Page 228: ...Discovery QDR Series Technical Manual 6 16...
Page 248: ......