Discovery QDR Series Technical Manual
9-6
9.1.3.3.4 Hi Gain.
This is a spinner control with a range of 0 to 11. Changing Hi Gain changes the displayed
detector values for the Hi portions of the current segment. The associated actual gain
values are shown in the following table.
9.1.3.3.5 Lo Gain.
This is a spinner control with a range of 0 to 11. Changing Lo Gain changes the displayed
detector values for the Lo portions of the current segment. The associated actual gain
values are shown in the above table.
9.1.3.3.6 Xray Mode.
This is a spinner control with a range of 0 to 8. The default mode is 3. This control
determines the voltage, current and duty cycle for the x-ray generator. Most scans use X-
ray Mode 3, Whole Body uses 3 and IVA (single energy) uses 6. The following table
shows the full set of defined X-ray modes.
Gain Code
A/D Gain
MUX Gain
Total Gain
0
1
2
1
1
1
2
2
2
1
4
4
3
1
8
8
4
2
1
2
5
2
2
4
6
2
4
8
7
2
8
16
8
4
1
4
9
4
2
8
10
4
4
16
11
4
8
32
X-Ray Mode
Pulse Cycle
Peak Ma
Average Ma
High KV
Low Kv
0
0
--
--
--
--
1
50%
3
0.75
140
100
2
50%
10
2.5
--
100
3
50%
10
240
100
4
50%
3
0.75
--
80
5
50%
3
0.75
140
80
6
100%
10
5.0
140
--
7
50%
10
2.5
140
--
8
100%
10
5.0
--
100
Summary of Contents for DISCOVERY A
Page 52: ...Discovery QDR Series Technical Manual 2 30...
Page 55: ...Discovery QDR Series Technical Manual 3 3 Figure 3 2 System Dimensions for Discovery C...
Page 56: ...Discovery QDR Series Technical Manual 3 4 Figure 3 3 System Dimensions for Discovery SL...
Page 144: ...Discovery QDR Series Technical Manual 4 22...
Page 188: ...Discovery QDR Series Technical Manual 5 22...
Page 228: ...Discovery QDR Series Technical Manual 6 16...
Page 248: ......