A-13
Internal driver error [E_INTERNAL_ERROR_I]
Comment: If returned in response to an ON_LINE command, indicates that the MAPI
device driver is not properly installed in the system.
Action: Verify that the line DEVICE=C:\MENU\MAPIDEV.SYS is in the system
CONFIG.SYS file, that the interrupt level (IRQ) and I/O base address are correctly set
and that the file C:\MENU|MAPIDEV.SYS is installed on the system.
Comment: If returned in response to a XRAY_ON_PHASE, XRAY_SEQUENCE, or
PULSE_PER_SEG command to the C-arm device, indicates that the currently selected
X-ray mode is invalid.
Action: Report this to Hologic Technical Support.
Comment: If reported as the device status for the C-arm, it indicates an unrecognized
response from the C-arm micro.
Action: This may be due to a communications error between the AT and the C-arm.
Follow the procedure under "BAD CHECKSUM IN MESSAGE." to diagnose I/O
errors.
Comment: If reported as the abort code for the state machine, it indicates that the
internal driver event queue was full when the state machine attempted to issue the
system-wide abort code for state machine termination.
Action: Report this error to Hologic Technical Support.
Internal Error:.....
Internal Software Error: ......
Software
Any message that begins this way should be reported to Hologic Technical Support.
Include as much information as possible about what you were doing at the time.
Internal stack error [E_STACK_ERROR_I]
Comment: Returned by the state machine driver to indicate that the local stack for a
task has overflowed or underflowed.
Action: Report this problem to Hologic.
Invalid ...
Software
In general, any message that begins with the word "Invalid" should be reported to
Hologic Technical Support. Include as much information as possible about what you
were doing at the time. See the following "Invalid..." errors for more information.
ERROR MESSAGE
HARDWARE
/SOFTWARE
ERROR
LOCUS
COMMENT/ACTION
Summary of Contents for DISCOVERY A
Page 52: ...Discovery QDR Series Technical Manual 2 30...
Page 55: ...Discovery QDR Series Technical Manual 3 3 Figure 3 2 System Dimensions for Discovery C...
Page 56: ...Discovery QDR Series Technical Manual 3 4 Figure 3 3 System Dimensions for Discovery SL...
Page 144: ...Discovery QDR Series Technical Manual 4 22...
Page 188: ...Discovery QDR Series Technical Manual 5 22...
Page 228: ...Discovery QDR Series Technical Manual 6 16...
Page 248: ......