Operation manual geodyna 2350 - 9413 126
Turning on the machine
28
29
30
31
32
E92 - Fig. 28
During the second attempt the distance gauge arm was still not
in the home position. The gauge arm is rendered inoperative.
D
Wait 3 seconds, or press the STOP key to continue.
E145 - Fig. 29
The contents of both permanent memories are different (but
both contain valid data).
Fatal error codes
The self-test program has detected an error and displays an al-
phanumeric code consisting of six digits and/or letters.
Note
The displayed error codes refer only to the internal line voltage
of the machine.
C10 800 - Fig. 30
Line voltage too low. Balancing is feasible if the motor can drive
the main shaft to the measuring speed. Wheel data may be lost.
D
Connect the machine to 110-125 V AC.
C10 801 - Fig. 31
Line voltage too high. Damage to the electronic unit of the ma-
chine is likely!
D
Turn off machine immediately.
Damage is not covered under warranty.
C10 804 - Fig. 32
Line voltage too high. Damage to the electronic unit of the ma-
chine is likely!
D
Turn off machine immediately.
Damage is not covered under warranty.