1.4.7
1 - 23
(2) Fix the cell holder by tightening the fixing screw on the cell
holder for reference.
Fig. 1-41 External View of Sample Compartment after Mounting
Cell Holder for Reference
1.4.7 Mounting 6 Cell Turret
(1) Check the drive motor shaft in the sample compartment. The
shaft has a slot (a hollow) thereon. Check then the axial
alignment portion (a protrusion) on the back of the 6 cell turret to
be coupled to the drive motor shaft. Attach the 6 cell turret to the
motor shaft so that the axial alignment portion of that turret will
fittingly mate to the slot on the drive motor shaft.
External view of the sample compartment
Shaft on back of the 6 cell turret
Fig. 1-42 Axial Alignment in Mounting 6 Cell Turret
Fixing screw
Drive motor shaft
for 6 cell turret
Slot
Screw
Axial alignment
portion