1.3.6
1 - 13
(6) The Auto Sipper is fixed on the main unit using guide screws.
Gently lift the Auto Sipper to check the guide screws on the
bottom. Also check the locating holes on the main unit sample
compartment.
Fig. 1-22 Guide Screws on Bottom of
Auto Sipper
Fig. 1-23 Locating Holes on Main Unit
Sample Compartment
(7) Place the Auto Sipper temporarily. In placing, provide a working
space behind the sipper to permit connecting the main unit
connection cables to the Auto Sipper. Place the Auto Sipper at a
position where the rear thereof comes approximately 10 to
15 cm ahead from the back of the sample compartment of the
photometer (that is the position where the sipper can sit stably).
Fig. 1-24 Temporary Placement of Auto Sipper
Guide screws
Auto Sipper
driving shaft
Locating holes
Provide space for connection
10 to 15 cm