1.4 Reverse Process to Change Auto Sipper to 6 Cell Turret
1 - 20
Loosen the hexagonal screw on the coupling shaft using the
hexagonal wrench which is an accessory to the spectrophotometer.
Screw loosening (side view)
Screw loosening (top view)
Fig. 1-35 Detaching Coupling Shaft (2)
1.4.4 Detaching Front Panel
Remove the front panel used on the Auto Sipper.
Slide upward to dismount.
Fig. 1-36 Detaching Front Panel
1.4.5 Attaching Sample Compartment Front Cover
In attaching the sample compartment front cover, check the tabs on
the front panel and on the instrument shown in Fig. 1-37. Engage the
left and right tabs on the sample compartment front cover
correspondingly with the tabs on the instrument, and then slide the
sample compartment front cover from the upper-side of the tab
downward (refer to Fig. 1-38).
Hexagonal
wrench
Hexagonal
wrench
Coupling
shaft