1.3.6
1 - 15
(10) To fix the Auto Sipper, tighten the guide screws using the slotted
screwdriver, which is an accessory to the spectrophotometer
main unit. After fixing the Auto Sipper, hold the upper side of the
front panel and the handle, and check if the sipper is firmly fixed
without looseness. If not fixed correctly, it indicates that the
sipper is placed on the skew. Redo placing the sipper in such
situation.
Fig. 1-26 Fixation with Guide Screws
NOTE:
Insufficient tightening of the guide screws causes
impediment in the Auto Sipper fixing conditions, in which
case there is a possibility of measurement accuracy
deterioration caused by optical axis misalignment of the
flow cell.
(11) Close the cover of the sample compartment.
Fig. 1-27 External View after Completion of Mounting of Auto Sipper
At this point, the mounting procedure of the Auto Sipper is
complete.
Guide screws