SAFETY - 6
SAFETY SUMMARY
NOTICE
Disposal of Waste Solution
Be sure to collect waste solution and treat it for proper disposal in
accordance with the relevant laws and regulations regarding water
pollution control and sewage treatment. Improper treatment of
waste solution may result in environmental pollution. Additionally, it
could lead to a penalty.
Accuracy and Precision of Measured Values
Carry out periodic inspection and check whether the system is
operating normally. Additionally, perform measurement of control
samples when necessary.
Electricity
Make sure that the power supply to the spectrophotometer is
100 V AC, 1 kVA or more (50 or 60 Hz).
Be sure to prepare a grounding wire together with power cables
and confirm that the wire has a ground resistance of 100
or less
(Class D grounding construction in Electric Facility Technical
Standards).
The high voltage circuit is used inside the spectrophotometer. To
prevent electric shock, do not open covers during operation unless
otherwise necessary.