1.4.2
1 - 17
Fig. 1-29 Holding Portions for Lifting Auto Sipper
(2) Place the Auto Sipper temporarily. In placing, provide a working
space behind the sipper to permit unplugging the main unit
connection cables from the Auto Sipper. Place the Auto Sipper
at a position where the rear thereof comes approximately 10 to
15 cm ahead from the back of the sample compartment of the
photometer (that is the position where the sipper can sit stably).
Fig. 1-30 Temporary Placement of Auto Sipper
NOTICE:
1. In handling the Auto Sipper, always hold the handle
and the upper portion of the front panel. Holding
other portion may cause damage or deviation in the
adjustment.
2. Avoid touching the flow cell while handling the Auto
Sipper. Touching the flow cell may cause deviation in
the optical axis alignment or damage of the cell.
Handle
Front panel
upper portion
Provide space for unplugging the connector
10 to 15 cm