2.2.4
2 - 7
2.2.4 Releasing Peristaltic Tube
Follow the procedure (1) through (5) above in reverse order, from (5)
to (1) (to detach the peristaltic tube from the roller).
NOTE:
The peristaltic tube must be released from the set position
while the Auto Sipper is not in use. Leaving the peristaltic
tube set on the roller leads to early deterioration of the tube.
2.3 Details of Operation Keys, etc.
Fig. 2-14
2.3.1 Sipping Lever and SIPPING Lamp
Pushing the sipping lever activates the peristaltic pump and turns on
the SIPPING lamp.
This lever is used for sipping liquid sample to be delivered to the flow
cell. After inserting the sipper nozzle into the liquid sample to be
sipped, pushing the sipping lever rotates the peristaltic pump for a
period of time set in advance, and the liquid sample is sipped and
delivered to the flow cell.
This lever can also be used together with the return key to return the
sample after completion of the measurement to a test tube. The
details of the procedure using the RETURN key are described in the
follwing section.
NOTE:
The sipping lever can be activated while the measurement
screen is displayed. It is not activated while the condition
setup screen is displayed.
RETURN key
WASH key
Sipping lever
Sipper nozzle
SIPPING lamp