1.3.5
1 - 7
NOTE:
If the front panel is not attached, the measurement will not
be correct because light from the outside of the instrument
may enter the sample compartment. The front panel cannot
be attached after the mounting of the Auto Sipper main unit;
the front panel must be attached at this installation step.
Fig. 1-10 Attaching Front Cover for Instrument
1.3.5 Attaching Coupling Shaft
Check the drive motor shaft in the sample compartment; the shaft
has a slot (a hollow) thereon. Check then the axial alignment portion
(a protrusion) on the lower end of the coupling shaft to be attached to
the drive motor shaft. Attach the coupling shaft to the drive motor
shaft so that the axial alignment portion of the coupling shaft will
fittingly mate to the slot on the drive motor shaft.
Fig. 1-11 Attaching Coupling Shaft
Hexagonal screw
Align tabs on the instrument
with tabs on the front panel
and slide the front panel from
the upper-side of the
instrument tab down to full
depth.
Appearance of sample compartment
Axial alignment
portion
Coupling Shaft
Insertion slot for
fixation
Drive motor
shaft for 6 cell
turret
Slot
Confirm that the front panel is
attached without level difference
with the instrument at the
portion indicated by
○
.