62
LAN Interface
Setting LAN communications
Set the commands shown below for the LAN communications.
:SYSTem:COMMunicate:LAN:IPADdress <Value 1>,<Value 2>,<Value 3>,<Value 4>
Set the IP address for the device.
:SYSTem:COMMunicate:LAN:SMASK <Value 1>,<Value 2>,<Value 3>,<Value 4>
Set the LAN subnet mask.
:SYSTem:COMMunicate:LAN:GATeway <Value 1>,<Value 2>,<Value 3>,<Value 4>
Set the address for the default gateway.
:SYSTem:COMMunicate:LAN:CONTrol <1 - 9999>
Specify the communication command port No.
:SYSTem:COMMunicate:LAN:UPDate
Update and apply the LAN settings.
When the communication setting mode switch (
DFLT
/
USER
) is set to
DFLT
, the default settings are
used regardless of the settings specified here.
Reference: "2.6 Setting the Communication Setting Mode" (p. 22)
Setting items
IP address
An address to identify individual instruments that are connected on the network. Be
sure to set an address that does not overlap with other devices.
Subnet mask
This setting divides the IP address into the address part showing the network and
the address part showing the device. Be sure to set a subnet mask that matches
the subnets of other devices within the same network.
Gateway
IP address
For network connection
When the communication PC and the device are in different networks, specify the
IP address of the device that becomes a gateway.
If the PC is in the same network, set a gateway address that is the same as the
default gateway in the general PC settings.
For one-to-one connection between the device and PC or when no gateway
is used
Set the IP address to
0.0.0.0
.
Communication
command port No.
Specify the TCP/IP port No. used for connecting communication commands.
HIOKI SW1001A961-04
Summary of Contents for SW9001
Page 2: ...HIOKI SW1001A961 04...
Page 42: ...38 Channel Delay Function HIOKI SW1001A961 04...
Page 48: ...44 Scan Measurement Example HIOKI SW1001A961 04...
Page 50: ...46 Checking the Device Status HIOKI SW1001A961 04...
Page 52: ...48 Initialization Settings HIOKI SW1001A961 04...
Page 118: ...114 Sample Programs 3 Select FILE Save All HIOKI SW1001A961 04...
Page 160: ...156 Outline Drawings 12 4 Outline Drawings SW1001 Unit mm Tolerance 0 2 HIOKI SW1001A961 04...
Page 161: ...157 Outline Drawings SW1002 Unit mm Tolerance 0 2 12 Appendix HIOKI SW1001A961 04...
Page 162: ...158 Outline Drawings HIOKI SW1001A961 04...
Page 164: ...160 HIOKI SW1001A961 04...
Page 167: ...HIOKI SW1001A961 04...
Page 168: ...HIOKI SW1001A961 04...
Page 169: ...HIOKI SW1001A961 04...