144
Accuracy Calculation Example
Example 2: Use of the device in combination with the BT4560
Used module
SW9002
Instrument
BT4560 RX function, 3 m
Ω
range, 1000 Hz, sampling SLOW
Measurement value
R 1 m
Ω
, X 0 m
Ω
R measurement accuracy = BT4560 R measurement ac SW9002 measurement influence
(BT4560 3 m
Ω
range R)
= [±(0.004 |R| + 0.0052 |X|) ± 8 dgt.] + (±0.1% f.s.)
= [±(0.004 × 1 m
Ω
+ 0.0052 × 0 m
Ω
) ± 0.0008 m
Ω
] + [±(0.1% ×
3.0000 m
Ω
)]
= (±0.0048 m
Ω
) + (±0.0030 m
Ω
)
= ±0.0078 m
Ω
(total error)
X measurement accuracy = BT4560 X measurement ac SW9002 measurement influence
(BT4560 3 m
Ω
range X)
= [±(0.004 |X| + 0.0052 |R|) ± 8 dgt.] + (±1.0% f.s.)
= [±(0.004 × 0 m
Ω
+ 0.0052 × 1 m
Ω
) ± 0.0008 m
Ω
] + [±(1.0% ×
3.0000 m
Ω
)]
= (±0.0052 m
Ω
) + (±0.0300 m
Ω
)
= ±0.0352 m
Ω
(total error)
Example 3: Use of the device in combination with the DM7276
Used module
SW9001
Instrument
DM7276 V function, 100 mV range, sampling MEDIUM
Measurement value
100 mV
Voltage measurement accuracy = DM7276 voltage measurement ac SW9001
measurement influence (all DM7276 voltage ranges)
= (±0.0015% rdg. ± 2 µV) + (±7 µV)
= (±0.0015% × 100 mV ± 2 µV) + (±7 µV)
= (±0.0035 mV) + (±7 µV)
= ±0.0105 mV (total error)
Example 4: Use of the device in combination with the RM3545
Used module
SW9001
Instrument
RM3545, LP OFF, 10 m
Ω
range, SPEED SLOW2, OVC ON
Measurement value
1 m
Ω
Resistance measurement accuracy = RM3545 resistance measurement ac SW9001
measurement influence (RM3545
10 m
Ω
range
, SPEED
SLOW2, OVC ON)
= (±0.060%
rdg. ± 0.001% f.s.) + (±0.006%
rdg. ± 0.001%
f.s.)
= (±0.060% × 1
m
Ω
± 0.001% × 10 m
Ω
) + (±0.006% × 1
m
Ω
± 0.001% × 10
m
Ω
)
= (±0.00060 m
Ω
±0.00010 m
Ω
) + (±0.00006 m
Ω
±0.00010 m
Ω
)
= ±0.00086
m
Ω
(total error)
HIOKI SW1001A961-04
Summary of Contents for SW9001
Page 2: ...HIOKI SW1001A961 04...
Page 42: ...38 Channel Delay Function HIOKI SW1001A961 04...
Page 48: ...44 Scan Measurement Example HIOKI SW1001A961 04...
Page 50: ...46 Checking the Device Status HIOKI SW1001A961 04...
Page 52: ...48 Initialization Settings HIOKI SW1001A961 04...
Page 118: ...114 Sample Programs 3 Select FILE Save All HIOKI SW1001A961 04...
Page 160: ...156 Outline Drawings 12 4 Outline Drawings SW1001 Unit mm Tolerance 0 2 HIOKI SW1001A961 04...
Page 161: ...157 Outline Drawings SW1002 Unit mm Tolerance 0 2 12 Appendix HIOKI SW1001A961 04...
Page 162: ...158 Outline Drawings HIOKI SW1001A961 04...
Page 164: ...160 HIOKI SW1001A961 04...
Page 167: ...HIOKI SW1001A961 04...
Page 168: ...HIOKI SW1001A961 04...
Page 169: ...HIOKI SW1001A961 04...