Connecting the Measurement Cable
2.4 Connecting the Measurement Cable
WARNING
Connect the measurement cable with the measuring object disconnected from
the measurement cable. The measuring object may be shorted depending on the
condition of the module switching circuit.
Please provide a suitable measurement cable.
See: “12.1 Measurement cable” (p. 145)
Measuring object
Tip of the probe
Disconnect
Disconnect
1
Turn off the device.
2
Disconnect the probe from the
measuring object.
Front
3
4
3
Connect the measurement
cable connector to the module’s
connector.
4
Secure the measurement cable
connector using the screws.
For information on connecting the measuring object, see the instruction manual of each instrument.