43
Scan Measurement Example
4.6 Scan Measurement Example
The following example shows scan measurement with the SW1001 and BT3562A connected using
EXT. I/O.
For acquiring measured values, the BT3562A data output function is used.
Wiring diagram
PC
SW1001
TRIG
BT3562A
EOM
USB
RS-232C
CLOSE
SCAN
BT3562A setting
External trigger
Set data output to on
SW1001 setting
Connection method: 4-wire
Scan list registration
SW1001
Start scanning
Scan input signal or
*TRG
command
BT3562A
Receive measured values.
Scan measurement
complete
Scan measurement start
Measured value
received for all
channels
Yes
No
4
Scan Function
HIOKI SW1001A961-04