95
Message Reference
Query Condition Register of Status Query Register Group
Syntax
Query
:STATus:QUEStionable:CONDition?
Response
<0 to 65535 (NR1)>
Description
Returns the present status query register value. Unused bits (indicated in the following
chart by “–”) return the value 0.
bit 15
bit 14
bit 13
bit 12
bit 11
bit 10
bit 9
bit 8
–
–
–
–
–
–
–
INFO_
ERR
bit 7
bit 6
bit 5
bit 4
bit 3
bit 2
bit 1
bit 0
BACKUP
_ERR
–
–
–
–
–
–
–
Example
:STAT:QUES:COND?
128
Backup error is occuring.
Note
•
The present status is returned when this command is received.
•Unlike
:STATus:QUEStionable:EVENt?
, :
STATus:QUEStionalbe?
, bit information
is not retained.
Query Event Register of Status Query Register Group
Syntax
Query
:STATus:QUEStionable[:EVENt]?
Response
<0 to 65535 (NR1)>
Description
Returns the retained status query register value. Unused bits (indicated in the following
chart by “–”) return the value 0.
bit 15
bit 14
bit 13
bit 12
bit 11
bit 10
bit 9
bit 8
–
–
–
–
–
–
–
INFO_
ERR
bit 7
bit 6
bit 5
bit 4
bit 3
bit 2
bit 1
bit 0
BACKUP
_ERR
–
–
–
–
–
–
–
Example
:STAT:QUES?
256
A hardware information error has occurred.
Note
•
When an event occurs, 1 is set for the bit corresponding to each event and retained.
•The bit is not cleared until
*CLS
or this query is executed or the power is turned on again.
Set and query Enable Register of Status Query Register Group
Syntax
Command
:STATus:QUEStionable:ENABle <0
–
65535(NR1)>
Query
:STATus:QUEStionable:ENABle?
Response
<0 to 65535 (NR1)>
Description
Sets the standard operation register mask pattern to a value from 0 to 65535. The value
when the device is turned on is 0. When issued as a query, unused bits are always set to 0.
bit 15
bit 14
bit 13
bit 12
bit 11
bit 10
bit 9
bit 8
–
–
–
–
–
–
–
INFO_
ERR
bit 7
bit 6
bit 5
bit 4
bit 3
bit 2
bit 1
bit 0
BACKUP
_ERR
–
–
–
–
–
–
–
Note
The initial value when the power is turned on is 0.
8
Communication Function
HIOKI SW1001A961-04
Summary of Contents for SW9001
Page 2: ...HIOKI SW1001A961 04...
Page 42: ...38 Channel Delay Function HIOKI SW1001A961 04...
Page 48: ...44 Scan Measurement Example HIOKI SW1001A961 04...
Page 50: ...46 Checking the Device Status HIOKI SW1001A961 04...
Page 52: ...48 Initialization Settings HIOKI SW1001A961 04...
Page 118: ...114 Sample Programs 3 Select FILE Save All HIOKI SW1001A961 04...
Page 160: ...156 Outline Drawings 12 4 Outline Drawings SW1001 Unit mm Tolerance 0 2 HIOKI SW1001A961 04...
Page 161: ...157 Outline Drawings SW1002 Unit mm Tolerance 0 2 12 Appendix HIOKI SW1001A961 04...
Page 162: ...158 Outline Drawings HIOKI SW1001A961 04...
Page 164: ...160 HIOKI SW1001A961 04...
Page 167: ...HIOKI SW1001A961 04...
Page 168: ...HIOKI SW1001A961 04...
Page 169: ...HIOKI SW1001A961 04...