104
Message Reference
Set and query IP address
Syntax
Command
:SYSTem:COMMunicate:LAN:IPADdress <Value 1>,<Value
2>,<Value 3>,<Value 4>
Query
:SYSTem:COMMunicate:LAN:IPADdress?
Response
<Value 1>,<Value 2>,<Value 3>,<Value 4>
<Value 1> = 0 to 255 (NR1)
<Value 2> = 0 to 255 (NR1)
<Value 3> = 0 to 255 (NR1)
<Value 4> = 0 to 255 (NR1)
<Value 1> to <Value 4> indicate the IP address.
Description
Set the IP address for the device.
The setting is enabled after
:SYSTem:COMMunicate:LAN:UPDate
(or restart) is executed.
When the communication setting mode switch (
DFLT
/
USER
) is set to
DFLT
, the address
is fixed to 192.168.0.254 regardless of the settings specified here.
The query returns the set value regardless of the communication setting mode switch state.
When the LAN function is set to off, 0,0,0,0 is returned.
Example
SYST:COMM:LAN:IPAD 192,168,0,2
SYST:COMM:LAN:UPD
SYST:COMM:LAN:IPAD?
192,168,0,2
Note
•
When connecting to the existing network, specify the LAN settings via USB or RS-232C
in advance. If a wrong IP address or overlapping IP address is used, a problem may
arise with the existing network.
•
An execution error occurs during scan operation (command).
Set and query subnet mask
Syntax
Command
:SYSTem:COMMunicate:LAN:SMASK <Value 1>,<Value 2>,
<Value 3>,<Value 4>
Query
:SYSTem:COMMunicate:LAN:SMASk?
Response
<Value 1>,<Value 2>,<Value 3>,<Value 4>
<Value 1> = 0 to 255 (NR1)
<Value 2> = 0 to 255 (NR1)
<Value 3> = 0 to 255 (NR1)
<Value 4> = 0 to 255 (NR1)
<Value 1> to <Value 4> indicate the mask value.
Description
Set the LAN subnet mask.
The setting is enabled after
:SYSTem:COMMunicate:LAN:UPDate
(or restart) is executed.
When the communication setting mode switch (
DFLT
/
USER
) is set to
DFLT
, the address
is fixed to 255.255.255.0 regardless of the settings specified here.
The query returns the set value regardless of the communication setting mode switch state.
Example
SYST:COMM:LAN:SMAS 255,255,255,0
SYST:COMM:LAN:UPD
SYST:COMM:LAN:IPAD?
255,255,255,0
Note
An execution error occurs during scan operation (command).
HIOKI SW1001A961-04
Summary of Contents for SW9001
Page 2: ...HIOKI SW1001A961 04...
Page 42: ...38 Channel Delay Function HIOKI SW1001A961 04...
Page 48: ...44 Scan Measurement Example HIOKI SW1001A961 04...
Page 50: ...46 Checking the Device Status HIOKI SW1001A961 04...
Page 52: ...48 Initialization Settings HIOKI SW1001A961 04...
Page 118: ...114 Sample Programs 3 Select FILE Save All HIOKI SW1001A961 04...
Page 160: ...156 Outline Drawings 12 4 Outline Drawings SW1001 Unit mm Tolerance 0 2 HIOKI SW1001A961 04...
Page 161: ...157 Outline Drawings SW1002 Unit mm Tolerance 0 2 12 Appendix HIOKI SW1001A961 04...
Page 162: ...158 Outline Drawings HIOKI SW1001A961 04...
Page 164: ...160 HIOKI SW1001A961 04...
Page 167: ...HIOKI SW1001A961 04...
Page 168: ...HIOKI SW1001A961 04...
Page 169: ...HIOKI SW1001A961 04...