34
Measurement between Two Instruments
3
Repeat channel selection and measurement using BT3562A for 8 channels (internal
resistance measurement).
4
Set the connection method to 2-wire.
The signal is connected to
TERMINAL 1
.
5
Repeat channel selection and measurement using DM7276 in the same way as for internal
resistance measurement (OCV measurement).
The sense wire for 4-wire CH1 to CH8 corresponds to 2-wire CH12 to CH19.
6
Set the connection method of
SLOT 2
to 2-wire.
Use
SLOT 2
for external potential for batteries since the measurement points are different.
The signal is connected to
TERMINAL 1
.
7
Repeat channel selection and measurement using DM7276 (external potential measurement
for batteries).
Control example using communication command
[SW1001] :SYST:MOD:WIRE:MODE 1,WIRE4
Set
SLOT 1
to 4-wire (connect to
TERMINAL 2
).
[SW1001] :CLOSE 101
Select
SLOT 1
, CH1.
[SW1001] *OPC?
Check that the channel relay has been closed.
[SW1001] 1
Receive a response “1” to the
*
OPC?
query.
[BT3562A] :READ?
Execute single measurement using BT3562A.
[BT3562A] 1.0258E-3
Receive measured values.
[SW1001] :CLOSE 102
Select the next CH2.
...
(Repeat until CH8.)
[SW1001] :SYST:MOD:WIRE:MODE 1,WIRE2
Set
SLOT 1
to 2-wire (connect to
TERMINAL 1
).
[SW1001] :CLOSE 112
Select
SLOT 1
, CH12 (4-wire CH1 sense).
[SW1001] *OPC?
Check that the channel relay has been closed.
[SW1001] 1
Receive a response “1” to the
*
OPC?
query.
[DM7276] :READ?
Execute single measurement using DM7276.
[DM7276] +03.00
Receive measured values.
[SW1001] :CLOSE 113
Select the next CH13 (4-wire CH2 sense).
...
(Repeat until CH19.)
[SW1001] :SYST:MOD:WIRE:MODE 2,WIRE2
Set
SLOT 2
to 2-wire (connect to
TERMINAL 1
).
[SW1001] :CLOSE 201
Select
SLOT 2
, CH1.
[SW1001] *OPC?
Check that the channel relay has been closed.
[SW1001] 1
Receive a response “1” to the
*
OPC?
query.
[DM7276] :READ?
Execute single measurement using DM7276.
[DM7276] +00.00
Receive measured values.
[SW1001] :CLOSE 202
Select the next CH2.
...
(Repeat until CH8.)
HIOKI SW1001A961-04
Summary of Contents for SW9001
Page 2: ...HIOKI SW1001A961 04...
Page 42: ...38 Channel Delay Function HIOKI SW1001A961 04...
Page 48: ...44 Scan Measurement Example HIOKI SW1001A961 04...
Page 50: ...46 Checking the Device Status HIOKI SW1001A961 04...
Page 52: ...48 Initialization Settings HIOKI SW1001A961 04...
Page 118: ...114 Sample Programs 3 Select FILE Save All HIOKI SW1001A961 04...
Page 160: ...156 Outline Drawings 12 4 Outline Drawings SW1001 Unit mm Tolerance 0 2 HIOKI SW1001A961 04...
Page 161: ...157 Outline Drawings SW1002 Unit mm Tolerance 0 2 12 Appendix HIOKI SW1001A961 04...
Page 162: ...158 Outline Drawings HIOKI SW1001A961 04...
Page 164: ...160 HIOKI SW1001A961 04...
Page 167: ...HIOKI SW1001A961 04...
Page 168: ...HIOKI SW1001A961 04...
Page 169: ...HIOKI SW1001A961 04...