91
Continuous Voltage Rise Test
Setting the number of tests
The number of tests is set.
In a BDV test, determine the dielectric breakdown strength or dielectric breakdown voltage from the
median of the result of the set number of tests.
1
1
1
Press
[BASIC]
>
[TEST COUNT]
on the setting
screen.
2
2
2
Change the value using
[▲]
or
[▼]
and then
press
[ENTER]
.
1
*
1
to
20
*1. Default setting
Setting the dielectric breakdown judgment reference value
The dielectric breakdown judgment method and reference value are set.
There are two judgment methods, judgment by the current upper-limit reference value and the arc
discharge detection.
(1) Setting the current upper-limit value
1
1
1
Press
[BASIC]
>
[UPPER]
on the setting screen.
2
2
2
Change the value using
[▲]
or
[▼]
and then
press
[ENTER]
.
0.010 mA
*
2
to
100 mA
*2. Default setting
Dielectric Breakdown V
oltage (BDV) Measurement Mode
HIOKI ST5680A961-00