228
Error Display
CONTACT SETTING
ERROR
There is an error in the setting
for a contact check.
−
Set the voltage drop time setting to off.
The compensation value of the
contact check has not been
acquired.
Execute an open compensation of the
contact check. (p. 56)
There is an error in the test
voltage setting. (compensation
execution screen only)
Set the test voltage to 100 V or more.
HARDWARE ERROR
An error in the internal circuit
has been detected.
−
The instrument is malfunctioning. Turn
off the instrument and make a repair
request.
Other errors
Error Display
Description
Ext. I/O output
Solution
PANEL LOAD FAILED
Panel load has failed.
−
Check the panels that have been saved.
ADJUST FAILED
Offset cancel (open
compensation) has failed.
−
The current may be over the current
range that can be compensated. Check
the connection and execute the offset
cancel again. (p. 57)
HARDWARE ERROR
(at start up)
An error in the internal circuit
has been detected.
−
The instrument is malfunctioning. Turn
off the instrument and make a repair
request.
HIOKI ST5680A961-00
Summary of Contents for ST5680
Page 2: ...HIOKI ST5680A961 00 ...
Page 84: ...84 Individually Executing Specific Steps HIOKI ST5680A961 00 ...
Page 118: ...118 Key Lock HIOKI ST5680A961 00 ...
Page 154: ...154 Calibration Expiration Check Function HIOKI ST5680A961 00 ...
Page 182: ...182 Internal Circuit Configuration HIOKI ST5680A961 00 ...
Page 220: ...220 Option Specifications HIOKI ST5680A961 00 ...
Page 238: ...238 External View 16 4 External View Unit mm HIOKI ST5680A961 00 ...
Page 250: ...250 HIOKI ST5680A961 00 ...
Page 253: ...www hioki com HIOKI ST5680A961 00 ...