194
Connecting and Setting RS-232C (Z3001)
Setting procedure
Set RS-232C communication settings using the system screen. Settable only when a Z3001 RS-
232C interface is mounted to the instrument.
2
2
3
3
4
4
5
5
1
Press
[SYS]
>
[I/F]
on the measurement screen.
2
Press
[RS-232C]
.
3
Press
[SPEED]
and then select the Baud rate.
9600
*
1
,
19200
,
38400
,
57600
*1. Default setting
4
Press
[HANDSHAKE]
and then select
handshake.
OFF
*
2
No flow control
SOFTWARE
Software control (XON/XOFF)
*2. Default setting
5
Select the terminator.
CRLF
*
3
,
CR
,
LF
*3. Default setting
HIOKI ST5680A961-00