Automatic Discharge Function
4.17 Automatic Discharge Function
This function automatically switches the discharge circuit to discharge the internal circuit after each
test.
DANGER
Until the
DANGER
lamp turns off, never touch the voltage output
terminals, test lead, and test target.
When a withstand voltage test or insulation resistance test is performed for the test
target containing capacitive components, these areas remain charged, which could
cause serious bodily injury.
This instrument releases residual electric charge after each test and discharges
these areas (Discharge resistance: 700 k
Ω
).
• The
DANGER
lamp is lit during discharge.
•
When the voltage between output terminals falls below the safe voltage (approx. 60 V DC), the
DANGER
lamp turns off.
•
The larger the capacitance is, the longer the discharge takes.