207
Functional Specifications
14.3
Functional Specifications
-1. Withstand voltage test mode
Judgment function
Judgment operation
Judgment
Judgment procedure
Display
Buzzer
External I/O
Upper fail
If
(measured value) > (upper reference
value)
is satisfied, the output is cut off, and
an upper-fail judgment is given. Judgment
is not performed during the judgment wait
time (delay time) beginning from the test
start.
Displayed on
the LCD
On
U_FAIL signal on
Lower fail
If
(measured value) < (lower reference
value)
is satisfied, the output is cut off, and
a lower-fail judgment is given. Judgment
not performed during the voltage rise time.
Displayed on
the LCD
On
L_FAIL signal on
Pass
When
(upper reference value) ≥ (measured
value) ≥ (lower reference value)
is satisfied
after the set time has elapsed, the output is
shut off, and a pass judgment is given.
Displayed on
the LCD
On
PASS signal on
•The PASS signal is continuously output until the next test starts.
•The UPPER FAIL and LOWER FAIL signals are continuously output until the next
test starts.
•The fail and pass buzzer volume can be adjusted.
Upper-limit reference
value setting
0.010 mA to 100.0 mA (default setting: 0.011 mA)
Lower-limit reference
value setting
0.010 mA to 100 mA, off (default setting: Off, 0.010 mA)
Set resolution
0.001 mA (0.010 mA to 9.999 mA)
0.01 mA (10 mA to 99.99 mA)
Judgment accuracy
For 1.000 mA or more:
±
(1.5% of setting)
For less than 1.000 mA:
±
(1.5% of s30 µA)
Automatic discharge
function
Operations
Residual electric charge is discharged through the internal
circuit at the completion of a DC withstand voltage test.
(discharge resistance: 700 k
Ω
)
Start voltage
Operations
Setting range
Default setting
The voltage at the start of a test can be set as a start
voltage.
0% to 99% of the test voltage (resolution: 1%)
0%
Limit voltage value
Operations
Setting range
Default setting
The upper-limit reference value for the test voltage setting
can be set to prevent excessive voltage from being applied
to the test target due to operation errors, etc.
0.010 kV to 8.000 kV
8.000 kV
Output voltage monitoring
function
Operations
The output is shut down and the test ends when the output
voltage exceeds
±
(5% of s50 V) after the voltage
rise time or the voltage deviates from the set test voltage
during the test and does not fall within
±
10 dgt of the set
voltage within 5 seconds.
Resistance value display
Operations
The resistance value, calculated by dividing the measured
voltage value by measured current value, of the test target
during a test is displayed.
Maximum current value
display
Operations
The maximum value of the measured current value during a
test can be displayed.
Specifications
HIOKI ST5680A961-00
Summary of Contents for ST5680
Page 2: ...HIOKI ST5680A961 00 ...
Page 84: ...84 Individually Executing Specific Steps HIOKI ST5680A961 00 ...
Page 118: ...118 Key Lock HIOKI ST5680A961 00 ...
Page 154: ...154 Calibration Expiration Check Function HIOKI ST5680A961 00 ...
Page 182: ...182 Internal Circuit Configuration HIOKI ST5680A961 00 ...
Page 220: ...220 Option Specifications HIOKI ST5680A961 00 ...
Page 238: ...238 External View 16 4 External View Unit mm HIOKI ST5680A961 00 ...
Page 250: ...250 HIOKI ST5680A961 00 ...
Page 253: ...www hioki com HIOKI ST5680A961 00 ...