160
External Input and Output Terminals and Signals
Signal functions
Input signal
START
Starts a test when it is on.
INTERLOCK
Releases the interlock and allows the instrument to be operated when it is
on. The interlock is activated and the instrument enters the protection state
when this signal is turned off. This signal is enabled regardless of the EXT_
EN signal status.
See “Interlock function” (p. 165).
LOAD0 to LOAD7
Allows you to select the test conditions saved using the panel save function.
STOP
Ends the test when it is on. This signal is enabled regardless of the EXT_EN
signal status.
EXT_EN
Enables the external I/O input signal when it is on. The INTERLOCK and
STOP signals are enabled regardless of the status of this signal.
LD_VALID
Executes panel load selected by LOAD0 - LOAD7 when it is on.
Output signal
ERR
Outputs on when a measurement system error occurs
U_FAIL
Outputs on when the instrument fails with a test upper-limit value error. The
output timing can be selected from two patterns.
L_FAIL
Outputs on when the instrument fails with a test lower-limit value error. The
output timing can be selected from two patterns.
H.V.ON
Outputs on when a voltage is applied and remains in the voltage output
terminals.
W-FAIL
Outputs on when the instrument fails a withstand voltage test.
W-MODE
Outputs on while a withstand voltage test is selected in the test mode.
STEP_END
Outputs on when one step is completed during a program test.
ARC_DET
Outputs on when the arc discharge is detected during a withstand voltage
test. (p. 92)
PASS
Outputs on during pass judgment. The output timing can be selected from
two patterns.
OUT0 to OUT1
General purpose output signals. On is output when the instrument enters the
set state.
TEST
Outputs on during a test time. A signal can be selected from four types with
varying output timing.
READY
Outputs on when the instrument enters the ready state.
PROTECTION
Outputs on when the protection function is activated.
CONT_ERR
Outputs on when a contact check error occurs.
IR-FAIL
Outputs on when the instrument fails an insulation resistance test.
IR-MODE
Outputs on while an insulation resistance test is selected in the test mode.
PROG_END
Outputs on when one program is completed during a program test.
HIOKI ST5680A961-00
Summary of Contents for ST5680
Page 2: ...HIOKI ST5680A961 00 ...
Page 84: ...84 Individually Executing Specific Steps HIOKI ST5680A961 00 ...
Page 118: ...118 Key Lock HIOKI ST5680A961 00 ...
Page 154: ...154 Calibration Expiration Check Function HIOKI ST5680A961 00 ...
Page 182: ...182 Internal Circuit Configuration HIOKI ST5680A961 00 ...
Page 220: ...220 Option Specifications HIOKI ST5680A961 00 ...
Page 238: ...238 External View 16 4 External View Unit mm HIOKI ST5680A961 00 ...
Page 250: ...250 HIOKI ST5680A961 00 ...
Page 253: ...www hioki com HIOKI ST5680A961 00 ...