161
External Input and Output Terminals and Signals
Setting the judgment signal output timing
•
The output timing for judgment output signals (PASS, U_FAIL, and L_FAIL) can be selected from
two patterns.
• The DCW and IR judgment output timing can be set individually.
• When the contact check function is
[ON]
, the output timing for judgment signals is
[MEAS END]
(p. 105) only.
DCW judgment output timing
1
1
1
Press
[SYSTEM]
>
[JUDGE W]
on the
system screen.
2
2
2
Select the setting.
MEAS END
*
1
Turns on the output as soon
as the test voltage is turned off
when a test is completed.
TEST END
Turns on the output as soon
as a test is completed.
*1. Default setting
IR judgment output timing
1
1
1
Press
[SYSTEM]
>
[JUDGE IR]
on the
system screen.
2
2
2
Select the setting.
MEAS END
*
2
Turns on the output as soon
as the test voltage is turned off
when a test is completed.
TEST END
Turns on the output as soon
as a test is completed.
*2. Default setting
External Control (EXT
. I/O)
HIOKI ST5680A961-00
Summary of Contents for ST5680
Page 2: ...HIOKI ST5680A961 00 ...
Page 84: ...84 Individually Executing Specific Steps HIOKI ST5680A961 00 ...
Page 118: ...118 Key Lock HIOKI ST5680A961 00 ...
Page 154: ...154 Calibration Expiration Check Function HIOKI ST5680A961 00 ...
Page 182: ...182 Internal Circuit Configuration HIOKI ST5680A961 00 ...
Page 220: ...220 Option Specifications HIOKI ST5680A961 00 ...
Page 238: ...238 External View 16 4 External View Unit mm HIOKI ST5680A961 00 ...
Page 250: ...250 HIOKI ST5680A961 00 ...
Page 253: ...www hioki com HIOKI ST5680A961 00 ...