
49
Setting the Test Time
IR setting
1
1
1
Press
[BASIC]
>
[TEST TIME]
on the setting
screen.
2
2
2
Change the value using
[▲]
or
[▼]
and then
press
[ENTER]
.
0.1 s
to
999 s
,
CONT
Default setting: 1.0 s
The lower limit is set by
[RISE TIME]
(voltage rise time)
and
[DELAY]
(judgment wait time).
[DELAY]
<
[RISE TIME]
+
[TEST TIME]
(Only when
[DELAY]
is not set to off)
Withstand V
oltage
Test/Insulation Resistance
Test Mode
HIOKI ST5680A961-00
Summary of Contents for ST5680
Page 2: ...HIOKI ST5680A961 00 ...
Page 84: ...84 Individually Executing Specific Steps HIOKI ST5680A961 00 ...
Page 118: ...118 Key Lock HIOKI ST5680A961 00 ...
Page 154: ...154 Calibration Expiration Check Function HIOKI ST5680A961 00 ...
Page 182: ...182 Internal Circuit Configuration HIOKI ST5680A961 00 ...
Page 220: ...220 Option Specifications HIOKI ST5680A961 00 ...
Page 238: ...238 External View 16 4 External View Unit mm HIOKI ST5680A961 00 ...
Page 250: ...250 HIOKI ST5680A961 00 ...
Page 253: ...www hioki com HIOKI ST5680A961 00 ...