Internal Circuit Configuration
12.4
Internal Circuit Configuration
NPN setting
30 V max.
5 V
Internal insulation common
(It is isolated from the protective ground of
the instrument.)
Internal isolated
power supply
Instrument
PLC, etc.
Output
Common
Input
Common
PLC, etc.
Zener voltage 30 V
Do not connect external power to pin 8.
PNP setting
5 V
Internal insulation common
(It is isolated from the protective ground of
the instrument.)
Internal isolated
power supply
Instrument
Output
Common
Input
Common
Do not connect external power to pin 8.
PLC, etc.
PLC, etc.
Zener voltage 30 V
30 V max.
Use ISO_COM for the common terminal for both the input and output signals.
External Control (EXT