161
ISO.COM
Internal GND
• External interface pin arrangement list (p. 84)
(3) TEST signal OFF timing setting
After a test, timing for the TEST signal output transistor to turn OFF from ON can be set.
FAST
The TEST signal turns OFF immediately after a test is completed
without waiting for the voltage between the measurement terminals to
drop.
SLOW
The TEST signal turns OFF after the voltage between the
measurement terminals drops below 10 V ± 3 V.
Refer to “Test duration accuracy” (p. 155).
EXT.SW (not isolated against the power supply protective ground terminal and
isolated against the measurement terminals)
The switch signal line plug of the 9299 Switched Probe is connected.
The operation mode can be switched between two modes, CONT and TRIG (the instrument
settings have priority).
Continuous (CONT)
mode
Measurement is performed only while the 9299 switch is being
pressed.
Trigger (TRIG) mode
A test starts and stops every time the 9299 switch is pressed.
RS-232C (not isolated against the power supply protective ground terminal and
isolated against the measurement terminals)
RMT
is displayed at the top of the screen during communications and any keys other than the
LOCAL key and STOP key are disabled.
Remote control and measured value output can be performed.
Output data setting: OFF/TYPE 1/TYPE 2
• TYPE1
Output: Measured value + judgment
• TYPE2
Output: Measured value (index display 000.0E+06)
(Same as
:MEASure?
(1) RS-232C interface
Transfer method
Communications: Full duplex
Synchronization: Start-stop synchronization
Baud rate
9,600 bps (default) / 19,200 bps / 38,400 bps
Data length
8 bits
Parity
None
Stop bit
1 bit
10
9
8
7
6
5
4
3
2
1
Appx.
Ind.
Summary of Contents for ST5520
Page 2: ......
Page 8: ...vi Contents ...
Page 22: ...14 Usage Notes ...
Page 56: ...48 Announcing the Judgment Results Using a Beep Sound ...
Page 82: ...74 Initializing the System Reset ...
Page 88: ...80 Deleting the Panel Data ...
Page 158: ...150 Command Reference ...
Page 193: ...Appx 13 Rack Mounting Screw location dimensional drawing 10 9 8 7 6 5 4 3 2 1 Appx Ind ...
Page 194: ...Appx 14 Dimensional Diagram Appx 8 Dimensional Diagram 166 2 215 1 14 1 80 1 ...
Page 197: ...13 09 ...
Page 198: ......
Page 199: ......
Page 200: ......
Page 201: ......