134
Command Reference
:MOHM:RANGe
Sets the resistance range.
Syntax
:MOHM:RANGe <data>
<data>
2 M / 20 M / 200 M / 2000 M / 4000 M / AUTO (text data)
Example
To set to 200 M
Ω
range
PC>
:MOHM:RANGe 200M
Error
An execution error occurs if the 2000 M
Ω
range is set when the test
voltage is 25 V to 99 V.
:MOHM:RANGe?
Queries the resistance range. The resistance range is returned in text data (<data>).
Syntax
:MOHM:RANGe?
Response
:MOHM:RANGE <data>
or simply
<data>
<data>
2 M / 20 M / 200 M / 2000 M / 4000 M / AUTO (text data)
Example
To query the resistance range (when the resistance range is 200 M
Ω
)
PC>
:MOHM:RANGe?
ST5520>
:MOHM:RANGE 200M
(when the
:HEADer
command is ON)
ST5520>
200M
(when the
:HEADer
command is OFF)
:SPEed
Sets the measurement speed to FAST/SLOW.
Syntax
:SPEed <data>
<data>
FAST/SLOW (text data)
Example
To set the measurement speed to FAST
PC>
:SPEed FAST
:SPEed?
Queries the measurement speed. The measurement speed setting FAST/SLOW is returned in
text data (<data>).
Syntax
:SPEed?
Response
SPEED <data>
or simply
<data>
<data>
FAST/SLOW (text data)
Example
To query the measurement speed (when the speed is set to FAST)
PC>
:SPEed?
ST5520>
:SPEED FAST
(when the
:HEADer
command is ON)
ST5520>
FAST
(when the
:HEADer
command is OFF)
Summary of Contents for ST5520
Page 2: ......
Page 8: ...vi Contents ...
Page 22: ...14 Usage Notes ...
Page 56: ...48 Announcing the Judgment Results Using a Beep Sound ...
Page 82: ...74 Initializing the System Reset ...
Page 88: ...80 Deleting the Panel Data ...
Page 158: ...150 Command Reference ...
Page 193: ...Appx 13 Rack Mounting Screw location dimensional drawing 10 9 8 7 6 5 4 3 2 1 Appx Ind ...
Page 194: ...Appx 14 Dimensional Diagram Appx 8 Dimensional Diagram 166 2 215 1 14 1 80 1 ...
Page 197: ...13 09 ...
Page 198: ......
Page 199: ......
Page 200: ......
Page 201: ......