95
Timing Chart
10 V
t1
t2
t3
t5
t4
/START
(Input)
/STOP
(Input)
Test voltage
Measured
value display
/PASS
(Output)
/U.FAIL
/L.FAIL
(Output)
/ERR
(Output)
t1
t2 t6
t5
/TEST
(Output)
Measurement end
When the TEST
signal OFF
timing setting is
SLOW
When measured
value-2 result is PASS
Previous measured value
Previous result
Previous result
Previous result
Measurement-1 contact check result
Measurement,
judgment
(including
contact check)
Measurement-1
Measurement-2
OFF
ON
OFF
OFF
0/0.0/0.00 M
Ω
Measured value-1
Measured value-2
Item
Time
t1
START, STOP signal pulse width
5 ms
MIN.
t2
START, STOP signal detection time
5 ms
MAX.
*
t3 Response time (DELAY)
AUTO, 5 ms to 999.9 s
t4 Measurement
time
Contact check: OFF
30 ms (FAST), 480 ms (SLOW)
Contact check: ON
80 ms (FAST), 480 ms (SLOW)
t5 Measurement
interval
Contact check: OFF
50 ms (FAST), 500 ms (SLOW)
Contact check: ON
100 ms (FAST), 500 ms (SLOW)
t6
Discharge time (time for the output voltage to drop
below 10 V)
20 ms
MAX.
(during pure resistance
measurement)
*
When the START signal is input after the test voltage is changed, a time of up to 500 ms is
added to detect the START signal.
10
9
8
7
6
5
4
3
2
1
Appx.
Ind.
Summary of Contents for ST5520
Page 2: ......
Page 8: ...vi Contents ...
Page 22: ...14 Usage Notes ...
Page 56: ...48 Announcing the Judgment Results Using a Beep Sound ...
Page 82: ...74 Initializing the System Reset ...
Page 88: ...80 Deleting the Panel Data ...
Page 158: ...150 Command Reference ...
Page 193: ...Appx 13 Rack Mounting Screw location dimensional drawing 10 9 8 7 6 5 4 3 2 1 Appx Ind ...
Page 194: ...Appx 14 Dimensional Diagram Appx 8 Dimensional Diagram 166 2 215 1 14 1 80 1 ...
Page 197: ...13 09 ...