93
Timing Chart
FAIL STOP mode timing chart
When the test duration setting (TIMER) is ON and the measurement is performed by the /START
signal input from EXT.I/O or when the test duration setting (TIMER) is ON and the measurement
is performed by the START key on the panel
Setting
Test duration (TIMER) Response time (DELAY)
TEST signal OFF timing
ON, 45 ms to 999.9 s
AUTO, 5 ms to 999.9 s
FAST
When the test duration (TIMER) is set to OFF, tests are performed continuously
until a FAIL judgment is made.
Measurement-1
Measurement-2
10 V
t1
t2
t6
t3
t5
t4
/START
(Input)
/TEST
(Output)
Test voltage
Measured
value display
/PASS
(Output)
/U.FAIL
/L.FAIL
(Output)
/ERR
(Output)
t7
Measurement end
When the TEST signal OFF
timing setting is SLOW
When measured
value-1 result is PASS
When measured
value-2 result is FAIL
Previous measured value
Previous result
Previous result
Previous result
OFF
ON
OFF
OFF
OFF
ON
OFF
0/0.0/0.00 M
Ω
Measured value-1
Measured value-2
Measurement-1 contact check result
Measurement,
judgment
(including
contact check)
10
9
8
7
6
5
4
3
2
1
Appx.
Ind.
Summary of Contents for ST5520
Page 2: ......
Page 8: ...vi Contents ...
Page 22: ...14 Usage Notes ...
Page 56: ...48 Announcing the Judgment Results Using a Beep Sound ...
Page 82: ...74 Initializing the System Reset ...
Page 88: ...80 Deleting the Panel Data ...
Page 158: ...150 Command Reference ...
Page 193: ...Appx 13 Rack Mounting Screw location dimensional drawing 10 9 8 7 6 5 4 3 2 1 Appx Ind ...
Page 194: ...Appx 14 Dimensional Diagram Appx 8 Dimensional Diagram 166 2 215 1 14 1 80 1 ...
Page 197: ...13 09 ...
Page 198: ......
Page 199: ......
Page 200: ......
Page 201: ......