125
Data Format Table
8.5 Data Format Table
Measured value
2 M
Ω
range
0.000E+06 to 4.000E+06
4 decimal point
20 M
Ω
range
1.90E+06 to 9.99E+06
3 decimal point
10.00E+06 to 40.00E+06
4 decimal point
200 M
Ω
range
19.0E+06 to 99.9E+06
3 decimal point
100.0E+06 to 400.0E+06
4 decimal point
2000 M
Ω
range
190E+06 to 999E+06
3 digits
1000E+06 to 4000E+06
4 digits
4000 M
Ω
range
190E+06 to 999E+06
3 digits
1000E+06 to 9990E+06
4 digits
Over Flow
9999E+06
4 digits
Under Flow
0000E+06
4 digits
Error
*1
0000E+10
4 digits
*1
When a contact check error, short circuit check error, or output voltage error occurs;
otherwise no data are found.
Comparator upper/lower limit values
2 M
Ω
or less
0.000E+06 to 2.000E+06
4 decimal point
20 M
Ω
or less
2.01E+06 to 9.99E+06
4 decimal point
10.00E+06 to 20.00E+06
4 decimal point
200 M
Ω
or less
20.1E+06 to 99.9E+06
4 decimal point
100.0E+06 to 200.0E+06
4 decimal point
Over 200 M
Ω
201E+06 to 999E+06
4 decimal point
1000E+06 to 9990E+06
4 digits
Test duration timer
0.045 s to 999.999 s (resolution 0.001 s)
Response time timer
0.005 s to 999.999 s (resolution 0.001 s)
10
9
8
7
6
5
4
3
2
1
Appx.
Ind.
Summary of Contents for ST5520
Page 2: ......
Page 8: ...vi Contents ...
Page 22: ...14 Usage Notes ...
Page 56: ...48 Announcing the Judgment Results Using a Beep Sound ...
Page 82: ...74 Initializing the System Reset ...
Page 88: ...80 Deleting the Panel Data ...
Page 158: ...150 Command Reference ...
Page 193: ...Appx 13 Rack Mounting Screw location dimensional drawing 10 9 8 7 6 5 4 3 2 1 Appx Ind ...
Page 194: ...Appx 14 Dimensional Diagram Appx 8 Dimensional Diagram 166 2 215 1 14 1 80 1 ...
Page 197: ...13 09 ...
Page 198: ......
Page 199: ......
Page 200: ......
Page 201: ......