3.8 Using D/A Output
93
Setting the rectifier for analog output
1
Press
.
2
Press
to display the settings
screen shown to the left.
3
Press
or
to select Std.
4
Press
or
to select the channel
for which you wish to set the rectifier.
Settings:
U1
→
2
→
3
→
I1
→
2
→
3
→
P1
→
2
→
3
→
0
…
(U: voltage; I: current; P: power; 1, 2, 3: chan-
nel number; 0: sum)
5
Press
or
to set the rectifier.
The rectifier lamp will light up to reflect the set-
ting location.
Settings:
AC+DC
→
AC+DC Umn
→
DC
→
AC
→
FND
…
6
Press
to exit the settings.
The instrument will return to the normal mea-
surement state.