3.7 External Control
86
Integrated value reset (RESET terminal)
Display hold (HOLD terminal)
• When integration is started with external control, it can only be stopped by external control. The integration
set time is ignored. If you attempt to stop integration with
, [
Err.11
] will be displayed.
• A delay (data update interval) of up to 200 msec. will occur between input of the integration start signal
and the actual start of integration.
• The
EXT
lamp will light up while integration triggered by external control is being performed.
• While integration is being performed (while the
RUN
lamp is lit up), integrated values cannot be reset.
Attempting to do so will cause [
Err.15
] to be displayed.
• A delay (date update interval) of up to 200 msec. will occur between input of the integration reset signal
and the actual integration reset.
5 V (open)
0 V (shorted)
Integrated values are reset during this interval.
At least 200 msec.
A delay (data update interval) of up to 200 msec. will occur between input of the hold signal and the actual
hold.
5 V (open)
0 V (shorted)
At least 200 msec.
The display is held during this interval.
Display hold cancel
Display hold start
Summary of Contents for PW3336-01
Page 2: ......
Page 6: ...Contents iv...
Page 16: ...Operating Precautions 10...
Page 32: ...1 4 Measurement Workflow 26...
Page 44: ...2 7 Turning Off the Instrument 38...
Page 181: ...5 6 Wiring Specifications 175 5 3 phase 4 wire 3P4W Current sensor...
Page 182: ...5 6 Wiring Specifications 176...
Page 199: ...Appendix 4 Dimensional Diagram A13 e g PW3337 03 Appendix 4 Dimensional Diagram Unit mm...
Page 200: ...Appendix 4 Dimensional Diagram A14...
Page 203: ......
Page 204: ......
Page 205: ......
Page 206: ......