3.10 Other Functions
112
This section describes how to initialize the instrument’s settings. Initializing the instrument (performing a sys-
tem reset) reverts settings to their default values at the time the device was shipped from the factory. The sys-
tem reset should be performed while the self-test is in progress after turning on the instrument (before the
display changes to the normal display).
Example: PW3337-03
3.10.4Initializing the Instrument (System Reset)
• Disconnect any voltage or current input from the instrument before performing the reset.
• The RS-232C communications speed, GP-IB address, and LAN-related settings will not be initialized.
1
Turn on the instrument.
2
While the product model and ver-
sion are being displayed, press
and then
.
(The SHIFT lamp will not light up.)
3
Perform zero-adjustment as you
would during normal operation.
If the system reset is being performed, the dis-
play will appear as shown below after the com-
pletion of zero-adjustment.
4
The System Reset screen will be
displayed, and settings will be
reverted to their default values
when the instrument was shipped
from the factory.
Summary of Contents for PW3336-01
Page 2: ......
Page 6: ...Contents iv...
Page 16: ...Operating Precautions 10...
Page 32: ...1 4 Measurement Workflow 26...
Page 44: ...2 7 Turning Off the Instrument 38...
Page 181: ...5 6 Wiring Specifications 175 5 3 phase 4 wire 3P4W Current sensor...
Page 182: ...5 6 Wiring Specifications 176...
Page 199: ...Appendix 4 Dimensional Diagram A13 e g PW3337 03 Appendix 4 Dimensional Diagram Unit mm...
Page 200: ...Appendix 4 Dimensional Diagram A14...
Page 203: ......
Page 204: ......
Page 205: ......
Page 206: ......