5.3 Measurement specifications
147
Voltage Crest Factor Measurement Specifications
(Ucf: Displayed as CF V on panel display)
Current Crest Factor Measurement Specifications
(Icf: Displayed as CF A on panel display)
Time Average Current Measurement Specifications
(T.AV I: Displayed as T.AV A on panel display)
Time Average Active Power Measurement Specifications
(T.AV P: Displayed as T.AV W on panel display)
Measurement method
Calculates values from display values once each display update interval for voltage and
voltage waveform peak values.
Effective measuring range
As per voltage and voltage waveform peak value effective measurement ranges.
Display range
1.0000 to 612.00 (no polarity)
Over-range display
Displays [o.r] when displaying [o.r] for the voltage waveform peak value and when the
display voltage is [o.r] or 0.
Invalid data display
Ucfsum is displayed as [- - - - -].
Measurement method
Calculates values from display values once each display update interval for current and
current waveform peak values.
Effective measuring range
As per current and current waveform peak value effective measurement ranges.
Display range
1.0000 to 612.00 (no polarity)
Over-range display
Displays [o.r] when displaying [o.r] for the current waveform peak value and when the
display current is [o.r] or 0.
Invalid data display
Icfsum is displayed as [- - - - -].
Measurement method
Calculates the average by dividing the current integrated value by the integration time.
Measurement accuracy
(Current measurement accuracy) +(±0.01%rdg. ±1dgt.)
Effective measuring range
As per the current integration effective measurement range.
Display range
0% to 612% of the range (no [o.r] display)
Polarity
Has polarity when using the DC rectifier.
Peak exceeded warning
The unit (A) will flash when the integrated value contains data that triggers a PEAK
OVER I warning.
Invalid data display
Displays [- - - - -] when using the AC or FND rectifier.
Measurement method
Calculates the average by dividing the active power integrated value by the integration
time.
Measurement accuracy
(Active power measurement accuracy) + (±0.01%dg. ±1dgt.)
Effective measuring range
As per the active power integration effective measurement range.
Display range
0% to 3,745.4% of the range (no [o.r] display)
Polarity
Yes
Peak exceeded warning
The unit (W) will flash when the integrated value contains data that triggers a PEAK
OVER U or PEAK OVER I warning.
Invalid data display
Displays [- - - - -] when using the AC or FND rectifier.
Summary of Contents for PW3336-01
Page 2: ......
Page 6: ...Contents iv...
Page 16: ...Operating Precautions 10...
Page 32: ...1 4 Measurement Workflow 26...
Page 44: ...2 7 Turning Off the Instrument 38...
Page 181: ...5 6 Wiring Specifications 175 5 3 phase 4 wire 3P4W Current sensor...
Page 182: ...5 6 Wiring Specifications 176...
Page 199: ...Appendix 4 Dimensional Diagram A13 e g PW3337 03 Appendix 4 Dimensional Diagram Unit mm...
Page 200: ...Appendix 4 Dimensional Diagram A14...
Page 203: ......
Page 204: ......
Page 205: ......
Page 206: ......