6.2 Error Indication
180
Over-range
This error is displayed when the voltage or current exceeds 140%
of the range (or 1,060.5 V when using the 1,000 V voltage range).
See:
3.11.2 When o.r (over-range) Is Displayed (p.115)
3.2.4 Selecting the Voltage and Current Ranges (p.48)
Scaling error
This error is displayed when the display range has been exceeded
due to the VT ratio or CT ratio setting. Change the VT or CT ratio.
See:
3.2.9 Setting the VT and CT Ratio (p.59)
External current sensor input
This error is displayed when the a 200 mA to 5 A range key is
pressed while using external current sensor input.
See:
3.9 Using a Current Sensor (p.102)
No measurement or measured
value not ready
This error is displayed when switching the wiring mode, rectifier,
or channel. When starting averaging, the [-----] display indicates
the measured value when the first average value has been calcu-
lated.
Error display
Description
Cause and Remedy
Summary of Contents for PW3336-01
Page 2: ......
Page 6: ...Contents iv...
Page 16: ...Operating Precautions 10...
Page 32: ...1 4 Measurement Workflow 26...
Page 44: ...2 7 Turning Off the Instrument 38...
Page 181: ...5 6 Wiring Specifications 175 5 3 phase 4 wire 3P4W Current sensor...
Page 182: ...5 6 Wiring Specifications 176...
Page 199: ...Appendix 4 Dimensional Diagram A13 e g PW3337 03 Appendix 4 Dimensional Diagram Unit mm...
Page 200: ...Appendix 4 Dimensional Diagram A14...
Page 203: ......
Page 204: ......
Page 205: ......
Page 206: ......