3.10 Other Functions
108
You can hold the display of all measured values by pressing the
HOLD
key (placing the instrument in the dis-
play hold state).
Activating display hold
Canceling the display hold state
The display hold will be unavailable in the following condition:
• During the
AVG
lamp is blinking
The display hold will become available when the
AVG
lamp that has been blinking lights up, which means
that an average data is determined. Measurement values that are not supposed to be averaged such as a
peak value will not be subject to the display hold until an average data is determined.
• Immediately after the setting such as the range is changed (while [
- - - - -
] is shown on the display)
The display hold will become available when a measurement value appears on the display on which [
- - - - -
]
has been shown.
3.10 Other Functions
3.10.1Fixing Display Values (Display Hold)
1
Press
.
2
The display of measured values will
be fixed when
is pressed, and
the
HOLD
lamp will light up.
1
Press
while the instrument is
in the display hold state.
2
The instrument will revert to the nor-
mal measurement state, and the
HOLD
lamp will turn off.
The following operations are not available when in the display hold state:
• Range switching
The instrument will display [
Err.16
] if any range key is pressed. (p.179)
Similarly, the range will not change during auto-ranging operation. The range will be fixed to the range at
the time the
HOLD
key was pressed.
• The flashing cursor cannot be moved to parameters that cannot be changed while in the hold state.
• To change a setting, press the
HOLD
key to cancel the display hold state (so that the
HOLD
lamp turns
off).
Summary of Contents for PW3336-01
Page 2: ......
Page 6: ...Contents iv...
Page 16: ...Operating Precautions 10...
Page 32: ...1 4 Measurement Workflow 26...
Page 44: ...2 7 Turning Off the Instrument 38...
Page 181: ...5 6 Wiring Specifications 175 5 3 phase 4 wire 3P4W Current sensor...
Page 182: ...5 6 Wiring Specifications 176...
Page 199: ...Appendix 4 Dimensional Diagram A13 e g PW3337 03 Appendix 4 Dimensional Diagram Unit mm...
Page 200: ...Appendix 4 Dimensional Diagram A14...
Page 203: ......