3.9 Using a Current Sensor
107
Using an external CT
Usage example
Configure the instrument as follows:
Current sensor type: TYPE1
CT ratio: 200 (current sensor’s CT ratio of 20) × (external CT’s CT ratio of 10)
Current measurement range: 10 kA (shown as 50 A range on the instrument’s panel)
The current measured value (display value) on the instrument will be [7,800 kA].
If energized components are exposed when connecting the current sensor, exercise
care not to touch them or the CT. Doing so may result in electric shock, bodily injury,
or a short-circuit accident.
When using an external CT, avoid open-circuiting the secondary winding. If current
flows through the primary when the secondary is open, high voltage across the sec-
ondary could present a dangerous hazard.
• The external CT’s phase different could introduce a significant error component into power measurement.
For more accurate power measurement, use a CT with a small phase error in the frequency band used by
the circuit.
• When using an external CT, ground the CT’s secondary negative terminal to ensure safety.
Measured current
7,800 A (7.8 kA)
Current sensor
9669 Clamp on Sensor (rated current: 1,000 A AC)
External CT
10:1
Summary of Contents for PW3336-01
Page 2: ......
Page 6: ...Contents iv...
Page 16: ...Operating Precautions 10...
Page 32: ...1 4 Measurement Workflow 26...
Page 44: ...2 7 Turning Off the Instrument 38...
Page 181: ...5 6 Wiring Specifications 175 5 3 phase 4 wire 3P4W Current sensor...
Page 182: ...5 6 Wiring Specifications 176...
Page 199: ...Appendix 4 Dimensional Diagram A13 e g PW3337 03 Appendix 4 Dimensional Diagram Unit mm...
Page 200: ...Appendix 4 Dimensional Diagram A14...
Page 203: ......
Page 204: ......
Page 205: ......
Page 206: ......