3.7 External Control
85
External control uses the EXT. CONTROL terminals.
External control terminals and description of control
The external control terminals are input terminals for controlling the instrument by means of either 0/5 V logic
signals or short/open contact signals.
External control signals are detected using the intervals illustrated in the following timing diagrams, but there
may be a delay in the display depending on the input signal frequency, synchronization signal, timeout, and
other settings.
Integration start/stop (START/STOP terminal)
3.7
External Control
Terminal name
Description
START/STOP
Starts/stops integration.
When a high (5 V or open) or low (0 V or shorted) signal is input to this terminal, integration starts. When the
signal changes from low to high, integration stops.
RESET
Resets integrated values.
When this terminal is set to low for at least 200 ms, integrated values are reset during that period.
HOLD
Holds the display when this terminal changes from high to low. The display hold is canceled when the terminal
changes from low to high.
GND
Connect to the external device's GND terminal.
3.7.1 External control terminal
5 V (open)
0 V (shorted)
At least 400 msec.
Integration is performed during this interval.
Integration start
Integration stop
Summary of Contents for PW3336-01
Page 2: ......
Page 6: ...Contents iv...
Page 16: ...Operating Precautions 10...
Page 32: ...1 4 Measurement Workflow 26...
Page 44: ...2 7 Turning Off the Instrument 38...
Page 181: ...5 6 Wiring Specifications 175 5 3 phase 4 wire 3P4W Current sensor...
Page 182: ...5 6 Wiring Specifications 176...
Page 199: ...Appendix 4 Dimensional Diagram A13 e g PW3337 03 Appendix 4 Dimensional Diagram Unit mm...
Page 200: ...Appendix 4 Dimensional Diagram A14...
Page 203: ......
Page 204: ......
Page 205: ......
Page 206: ......