1.3 Names and Functions of Parts
19
Warning (WARNING) indicator lamps
Setting screen indicator lamps
These lamps indicate channel warnings. When the following warning lamps and er-
ror lamps are lit up, they indicate a hazardous state or inability to perform measure-
ment accurately:
PEAK OVER U
Lights up when an overvoltage input warning occurs, indi-
cating that the input voltage peak value has exceeded
±1,500 V or ±600% of the voltage measurement range.
PEAK OVER I
Lights up when an overcurrent input warning occurs, indi-
cating that the input current peak value has exceeded ±100
A or ±600% of the current measurement range.
SYNC. ERROR
Lights up when a synchronization error occurs, indicating
that synchronization cannot be detected.
HRM ERROR
Lights up when a harmonic measurement synchronization
error occurs, indicating that the harmonic measurement
synchronization frequency range was exceeded.
These lamps light up in response to settings on the following setting screens:
CH SET
Channel settings
• Synchronization source
• Current input method
• CT ratio
• VT ratio
• Zero-cross and frequency measurement filter
• Synchronization detection timeout
MEAS. SET
Shared settings that apply to all channels
• Integration time
• Number of averaging iterations
• Harmonic analysis upper limit order
• Synchronized measurement I/O (master, slave)
• D/A output
INTERFACE
Interface settings
• RS-232C
• GP-IB
• LAN
Summary of Contents for PW3336-01
Page 2: ......
Page 6: ...Contents iv...
Page 16: ...Operating Precautions 10...
Page 32: ...1 4 Measurement Workflow 26...
Page 44: ...2 7 Turning Off the Instrument 38...
Page 181: ...5 6 Wiring Specifications 175 5 3 phase 4 wire 3P4W Current sensor...
Page 182: ...5 6 Wiring Specifications 176...
Page 199: ...Appendix 4 Dimensional Diagram A13 e g PW3337 03 Appendix 4 Dimensional Diagram Unit mm...
Page 200: ...Appendix 4 Dimensional Diagram A14...
Page 203: ......
Page 204: ......
Page 205: ......
Page 206: ......